High-accuracy and cost-effective photodiode spectral response measurement system

dc.contributor國立臺灣師範大學機電工程學系zh_tw
dc.contributor.authorChang, Gao-Weien_US
dc.contributor.authorLin, Yu-Hsuanen_US
dc.contributor.authorYeh, Zong-Muen_US
dc.date.accessioned2014-10-30T09:36:13Z
dc.date.available2014-10-30T09:36:13Z
dc.date.issued2007-02-09zh_TW
dc.description.abstractWith the rapid growth of optoelectronics technologies, photodiodes (PDs) has been widely used in optical measurement systems, color measurement and analysis systems, etc. To meet most of the measurement requirements, the determination of PD spectral responses is very important. The goal of this paper is to develop a high-accuracy and cost-effective spectral response measurement system for PDs. In this paper, the proposed system contains a grating-based spectral filtering module, an amplifier module, and a digital-signal-processing (DSP) based platform. In the spectral filtering module, a single-grating monochromator based on a Czerny-Turner configuration is first analyzed and simulated, and then the experiments are conducted to check if the measurement accuracy is satisfactory. In the measurement system, optoelectronic signals from the PD under test are acquired from the amplifier module and the DSP-based platform is developed to communicate and manipulate the measured data. Through comparison with the measurement data from a commercially available system, it is found that our approach gives quite satisfactory results.en_US
dc.identifierntnulib_tp_E0402_02_031zh_TW
dc.identifier.issn0277-786Xzh_TW
dc.identifier.urihttp://rportal.lib.ntnu.edu.tw/handle/20.500.12235/36922
dc.languageenzh_TW
dc.publisherSPIEen_US
dc.relationaccepted by SPIE Symposium on Integrated Optoelectronic Devices 2007.(Paper Number 6471B-44)en_US
dc.relation.urihttp://dx.doi.org/10.1117/12.701656zh_TW
dc.titleHigh-accuracy and cost-effective photodiode spectral response measurement systemen_US